Feature Overview
The multi-channel calibrator comes standard with three output modules and one measurement module;
The output module has outputs of V, mV, mA, Ω Hz、 Signals from pulses, simulated thermocouples, and simulated thermal resistors;
The measurement module is capable of measuring V, mA, Ω Hz、 Signals from pulses, thermocouples, and thermal resistors;
Can trigger three output signals simultaneously;
Equipped with a stopwatch timing function, all three output signals can automatically start or stop timing according to requirements, with a timing range of 0.001~9999.999 seconds, solving the problem of time traceability during the calibration process and complying with the requirements of the "JJG 1003-2016 Flow Accumulator" calibration regulations;
The verification software of the flow accumulator can automatically check the density value of the medium, calculate the standard values of instantaneous flow rate and cumulative flow rate; Automatically fill data in the list, generate calibration reports, and store historical calibration data;
Support programmable ramp output, calibrate switch instruments, and automatically capture switch actions;
Support programmable step level output, with 8 commonly used step levels pre-set, each with 15 sets of data;
Support TC dedicated interface, Pt100 external sensor or constant temperature for cold end compensation;
Programmable unit conversion function, which can convert V, mA output or measurement into other engineering units;
When outputting mA signals, it supports two modes: analog transmitter and mA source;
Automatic identification of output/measurement ports and automatic indication of output/measurement range;
When outputting and measuring thermocouple ℃ signals, display the corresponding mV value;
When outputting and measuring the thermal resistance ℃ signal, display the corresponding Ω value;
When outputting and measuring Ω signals, it supports two, three, and four wire modes;
When outputting Hz signals, the amplitude can be set and supports square wave or sine wave modes;
Can set pulse frequency output and calibrate flow integrator instruments;
Provide 24V DC circuit power supply and measure the mA signal in the circuit;
Support signal fine-tuning on any digit of the output value, used for calibration of pointer instruments;
Can lock or unlock measurement values during measurement;
When measuring numerical values, collect the minimum and maximum values and calculate the average value;
Can be used as a high-precision platinum resistance digital thermometer and correct R0, a, b, c;
Support switch on/off testing;
Support pulse counting;
The accuracy can be calibrated by oneself without the need for factory repair.
Accurate and stable
Original 24K gold-plated low thermoelectric potential material, with more stable signal;
Timed output, better than 0.01 seconds.
Strong load capacity
10V load capacity: 10mA;
100mV load capacity: 10mA;
20mA load capacity: 1k Ω;
24V DC circuit power load capacity: 200mA;
50000Hz load capacity: 10k Ω.
safe and reliable
The output port can withstand 36V DC ingress;
Both the output port and the 24V DC port are capable of short-circuit self recovery.
Key Metrics
Output (one year, ambient temperature 20 ℃± 5 ℃) ★ represents the basic range gear
| Output range | Maximum allowable error of 0.01 level | resolution |
| 54.9999 V | ± (0.006% reading+0.0005 V) | 0.1 mV |
| ± 10.99999 V | ± (0.006% reading+0.0001 V) | 0.01 mV |
| ± 1.099999 V ★ | ± (0.006% reading+0.00001 V) | 1 μV |
| ± 109.9999 mV | ± (0.006% reading+0.002 mV) | 1 μV |
| 10 types of thermocouples ℃ | See "TC Thermocouple Technical Specifications" for details | 0.01 ℃ |
| 200.0000 mA | ± (0.012% reading+0.002 mA) | 0.1 μA |
| 20.00000 mA ★ | ± (0.006% reading+0.0002 mA) | 0.01 μA |
| 2.000000 mA | ± (0.006% reading+0.00008 mA) | 0.01 μA |
| 5000.00 Ω | ± (0.006% reading+0.05 Ω) | 10 mΩ |
| 500,000 Ω ★ | ± (0.006% reading+0.005 Ω) | 1 mΩ |
| 50.0000 Ω | ± (0.006% reading+0.003 Ω) | 1 mΩ |
| 8 types of thermal resistors ℃ | Please refer to the "RTD Thermistor Technical Specifications" for details | 0.01 ℃ |
| 54999.9 Hz ★ | ± 2 Hz | 0.1 Hz |
| 5499.99 Hz | ± 0.2 Hz | 0.01 Hz |
| 549.999 Hz | ± 0.02 Hz | 0.001 Hz |
| 999999 pulses | ± 1 piece | 1 piece |
| 24V DC circuit power supply | ± 0.2 V (when outputting 200 mA) |
Measurement (one year, ambient temperature 20 ℃± 5 ℃) ★ represents the basic range
| Measurement range | Maximum allowable error of 0.01 level | resolution |
| ± 109.9999 V | ± (0.006% reading+0.001 V) | 0.1 mV |
| ± 10.99999 | ± (0.006% reading+0.0001 V) | 0.01 mV |
| ± 1.099999 V ★ | ± (0.006% reading+0.00001V) | 1 μV |
| ± 109.9999 mV | ± (0.006% reading+0.002 mV) | 1 μV |
| 10 types of thermocouples ℃ | See "TC Thermocouple Technical Specifications" for details | 0.01 ℃ |
| 200.0000 mA | ± (0.012% reading+0.002 mA) | 0.1 μA |
| 20.00000 mA ★ | ± (0.006% reading+0.0002 mA) | 0.01 μA |
| 2.000000 mA | ± (0.006% reading+0.00008 mA) | 0.01 μA |
| 5499.99 Ω | ± (0.006% reading+0.05 Ω) | 10 mΩ |
| 549.999 Ω ★ | ± (0.006% reading+0.005 Ω) | 1 mΩ |
| 54.9999 Ω | ± (0.006% reading+0.003 Ω) | 1 mΩ |
| 8 types of thermal resistors ℃ | Please refer to the "RTD Thermistor Technical Specifications" for details | 0.01 ℃ |
| 59999.9 Hz ★ | ± 2 Hz | 0.1 Hz |
| 9999.99 Hz | ± 0.4 Hz | 0.01 Hz |
| 999.999 Hz | ± 0.04 Hz | 0.001 Hz |
| 999999 pulses | ± 1 piece | 1 piece |
Note: ① Preheating time: 30 minutes;
② Temperature coefficient: ± (0.0002% reading+0.0002% range)/℃ (<15 ℃ or>25 ℃);
③ Resistance and thermistor output excitation current range: 5000 Ω/(0.065mA~1.05mA);500/(0.65mA~10.5mA) ; 50Ω/(6.5 mA~10.5mA);
④ The pulse amplitude voltage is adjustable from 0.03V to 9.99V; The pulse frequency is adjustable from 1 to 50000 Hz.
RTD thermistor (one year, ambient temperature: 20 ℃± 5 ℃)
| Thermal resistance index number | temperature range | Maximum allowable error (analog output and measurement) |
| 0.01 level | ||
| Pt10 | (– 200~0)℃ | ± 0.11℃ |
| (0~400)℃ | ± 0.14℃ | |
| (400~600)℃ | ± 0.17℃ | |
| (600~850)℃ | ± 0.20℃ | |
| Pt100 | (– 200~0)℃ | ± 0.04℃ |
| (0~400)℃ | ± 0.07℃ | |
| (400~600)℃ | ± 0.09℃ | |
| (600~850)℃ | ± 0.11℃ | |
| Pt1000 | (– 200~0)℃ | ± 0.04℃ |
| (0~400)℃ | ± 0.07℃ | |
| (400~600)℃ | ± 0.09℃ | |
| (600~850)℃ | ± 0.11℃ | |
| Cu50 | (– 50~150)℃ | ± 0.05℃ |
| BA1 | (– 200~0)℃ | ± 0.05℃ |
| (0~400)℃ | ± 0.09℃ | |
| (400~650)℃ | ± 0.11℃ | |
| BA2 | (– 200~0)℃ | ± 0.04℃ |
| (0~400)℃ | ± 0.07℃ | |
| (400~650)℃ | ± 0.09℃ | |
| BA3 | (– 50~100)℃ | ± 0.04℃ |
| G | (– 50~150)℃ | ± 0.06℃ |
Note: ①comply withThe 90 international temperature scale is determined by the maximum allowable error of resistance output and measurement.
TC thermocouple (one year, ambient temperature: 20 ℃± 5 ℃)
| Thermocouple division number | temperature range | Maximum allowable error (analog output and measurement) |
| 0.01 level | ||
| S | (– 20~100)℃ | ± 0.61℃ |
| (100~400)℃ | ± 0.42℃ | |
| (400~1600)℃ | ± 0.34℃ | |
| (1600~1768)℃ | ± 0.36℃ | |
| R | (– 20~100)℃ | ± 0.61℃ |
| (100~600)℃ | ± 0.41℃ | |
| (600~1600)℃ | ± 0.30℃ | |
| (1600~1768)℃ | ± 0.36℃ | |
| J | (– 210~– 100)℃ | ± 0.19℃ |
| (– 100~400)℃ | ± 0.08℃ | |
| (400~900)℃ | ± 0.10℃ | |
| (900~1200)℃ | ± 0.13℃ | |
| K | (– 250~–150)℃ | ± 0.70℃ |
| (–150~1000)℃ | ± 0.15℃ | |
| (1000~1372)℃ | ± 0.19℃ | |
| T | (– 250~–200)℃ | ± 0.54℃ |
| (–200~0)℃ | ± 0.22℃ | |
| (0~400)℃ | ± 0.08℃ | |
| N | (– 200~– 100)℃ | ± 0.33℃ |
| (–100~800)℃ | ± 0.12℃ | |
| (800~1300)℃ | ± 0.17℃ | |
| E | (– 250~–100)℃ | ± 0.38℃ |
| (–100~700)℃ | ± 0.08℃ | |
| (700~1000)℃ | ± 0.11℃ | |
| B | (600~800)℃ | ± 0.56℃ |
| (800~1100)℃ | ± 0.42℃ | |
| (1100~1820)℃ | ± 0.35℃ | |
| WRe3-25 | (0~1800)℃ | ± 0.32℃ |
| (1800~2200)℃ | ± 0.46℃ | |
| (2200~2315)℃ | ± 0.60℃ | |
| WRe5-26 | (0~1000)℃ | ± 0.23℃ |
| (1000~1800)℃ | ± 0.36℃ | |
| (1800~2315)℃ | ± 0.59℃ |
Note: ①comply withThe 90 international temperature scale is determined by the maximum allowable error of mV output and measurement;
②Excluding sensor and compensation wire errors;
③Accuracy with external cold contacts, for internal contacts added0.2℃;
④The maximum allowable error of Pt100 temperature probe is ± 0.20 ℃.
Other indicators
| project | Specifications |
| Resolution of Pt100 temperature probe | 0.01℃ |
| Maximum allowable error of Pt100 temperature probe | ± 0.2℃ |
| Operating Temperature | 0℃~50℃ |
| Storage temperature | (-20~60)°C |
| Relative humidity (maximum, no condensation) | At 90% and 35 ℃; At 75% and 40 ℃; 45% at 50 ℃. |