Product Details
Feature Introduction
·support andJY7000The digital pressure module is used in conjunction to complete pressure calibration and precision measurement;
·Support simultaneous simulation and measurement of thermocouples, provide three cold end temperature compensation methods, and use an external Pt100 temperature probe for real-time temperature compensation; Using fixed temperature values for constant temperature compensation; Using built-in temperature sensors for cold end temperature compensation;
·When simulating and measuring thermocouples and thermal resistors, display the corresponding mV/Ω electrical value;
·Conversion: Supports output/measurement conversion to other data, supports linear, square, and square root conversion;
·Measurement record: Real time recording of measurement values and the ability to generate curve graphs;
·System settings: Supports network connections, and the system provides multiple software tools such as thermal engineering manuals, calculators, and unit conversions;
·Battery management: supports low battery prompt, charging prompt, battery status information display, etc., can be used while charging;
·7-inch large color industrial touch screen, no button design, can be directly operated by touch screen;
·Fashionable and high-end stress relieving straps that are both convenient to carry and free up hands during use;
·The range will be displayed on the screen, and there is no need to consult the user manual;
·Quick output: Save frequently used process calibration points, and each channel and signal can be output by pressing shortcut keys (long press to edit);
·The invention patent of "Numerical Control Variable Resistance Device" enables the calibrator to respond faster, more stably, and allow a wider excitation current when outputting resistance signals. Supports fast pulse RTD temperature transmitters andPLCCalibration;
·When outputting mA signals, two power supply methods are provided: internal and external;
·The frequency output supports both square waves and sine waves. The duty cycle and amplitude can be set for square waves, while the amplitude can be set for sine waves;
·When measuring mA signals, it can provide loop power to the transmitter;
·External 24V DC power supply, capable of supplying power to other devices;
·Switching of pressure and temperature units, mutual calculation of voltage, resistance, and temperature values, in compliance with the ITS-90 international temperature scale;
·Step: can set step level output signal;
·Slope: can set slope output signal, automatically calibrate switch instruments;
·Output and measurement have data conversion functions, and each channel and signal are independently designed;
·When measuring resistance and thermal resistance, support two -, three -, and four wire wiring methods;
·Equipped with conventional multimeter functions, it can measure signals such as frequency, diodes, switches, AC/DC voltage, capacitance, etc;
·Equipped with professional RCP63 measurement function, it can perform manual or automatic detection and calculation of thermocouple performance;
·Equipped with 50V or100VThe insulation resistance measurement function.
technical indicators
Process output (one year, ambient temperature)20℃±5℃)
Output range |
accuracy |
resolution |
internal resistance |
noise |
20 V |
0.01% reading+0.005% full degree |
1 mV |
≤0.2Ω |
≤20uV |
10 V |
0.008% reading+0.001% full degree |
100 uV |
||
1 V |
10 uV |
|||
100 mV |
0.008% reading+5uV |
1 uV |
≤50mΩ |
≤2uV |
20 mA |
0.01% reading+1uA |
1 uA |
≥20MΩ |
≤0.1uA |
4000 Ω |
0.008% reading+0.001% full degree |
100 mΩ |
External excitation current: 0.1mA |
|
400 Ω |
10 mΩ |
External excitation current: 1mA |
||
50000 Hz |
0.004% full degree |
1 Hz |
Square wave and sine wave; |
|
5000 Hz |
0.1 Hz |
|||
500 Hz |
0.01 Hz |
|||
Process measurement (one year, ambient temperature)20℃±5℃)
Measurement range |
accuracy |
resolution |
internal resistance |
10 V |
0.008% reading+0.001% full degree |
0.1 mV |
1 MΩ |
20 mA |
0.01% reading+1uA |
1 uA |
10 Ω |
Temperature probe measurement (one year, ambient temperature)20℃±5℃)
Measurement range |
accuracy |
resolution |
internal resistance |
±100 mV |
0.008% reading+5uV |
1 uV |
≥500 MΩ |
500 Ω |
0.008% reading+0.001% full degree |
10 mΩ |
/ |
5000 Ω |
100 mΩ |
/ |
Insulation testing (one year, ambient temperature)20℃±5℃)
output voltage |
display range |
resolution |
accuracy |
50 V/100 V |
0~9.99 MΩ |
0.01 M |
± (10% reading+5 words) |
10.0~99.9 MΩ |
0.1 M |
||
100~200 MΩ |
1 M |
Multimeter module (one year, ambient temperature)20℃±5℃)
Measurement range |
range |
accuracy |
capacitance |
2.000 uF |
± (4% reading+50 words) |
20.00 uF | ||
200.0 uF | ||
AC voltage |
5.000 V |
± (1% reading+50 words) |
50.00 V | ||
500.0 V | ||
DC voltage |
5.000 V |
± (0.1% reading+5 words) |
50.00 V | ||
500.0 V | ||
diode |
√ |
/ |
discrete signal |
On/off or level discrimination |
/ |
frequency |
50000 Hz |
0.004% full degree |
9999.9 Hz | ||
999.99 Hz |
RTD thermistor (one year, ambient temperature 20 ℃± 5 ℃)
Output/Measurement |
temperature range |
accuracy |
Pt100 |
(-200~850)℃ |
0.12 ℃ |
Pt1000 |
(-200~850)℃ |
0.12 ℃ |
Cu50 |
(-40~150)℃ |
0.05 ℃ |
G |
(-40~150)℃ |
0.05 ℃ |
BA1 |
(-190~650)℃ |
0.12 ℃ |
BA3 |
(-40~100)℃ |
0.06 ℃ |
TC thermocouple (one year, ambient temperature: 20 ° C)℃±5℃)
Output/Measurement |
range |
accuracy |
S |
(-50~0)℃ |
1.27℃ |
(0~100)℃ |
0.93℃ |
|
(100~1768)℃ |
0.68℃ |
|
R |
(-50~0)℃ |
1.36℃ |
(0~200)℃ |
0.95℃ |
|
(200~1768)℃ |
0.58℃ |
|
B |
(600~800)℃ |
0.92℃ |
(800~1820)℃ |
0.69℃ |
|
K |
(-250~-200)℃ |
1.14℃ |
(-200~-100)℃ |
0.36℃ |
|
(-100~600)℃ |
0.18℃ |
|
(600~1372)℃ |
0.28℃ |
|
N |
(-200~-100)℃ |
0.54℃ |
(-100~1300)℃ |
0.25℃ |
|
E |
(-250~-200)℃ |
0.60℃ |
(-200~-100)℃ |
0.23℃ |
|
(-100~700)℃ |
0.12℃ |
|
(700~1000)℃ |
0.15℃ |
|
J |
(-200~-100)℃ |
0.30℃ |
(-100~700)℃ |
0.13℃ |
|
(700~1200)℃ |
0.19℃ |
|
T |
(-250~-100)℃ |
0.87℃ |
(-100~0)℃ |
0.19℃ |
|
(0~400)℃ |
0.13℃ |
|
WRe3-25 |
(0-2315)℃ |
0.91℃ |
WRe3-26 |
(0-2315)℃ |
0.89℃ |
Other indicators
project |
Specifications |
Process simulation quantity temperature influence quantity |
<(10ppm reading+5ppm quantity) |
Verification table working environment |
0-45℃/0-90%rh |
Calibration table storage temperature |
-20~55℃ |
Voltage output carrying capacity |
>10mA, Output overcurrent and short circuit protection |
power supply |
DC 24V/1.6A |
battery |
14.4V 6000mAh/86.4Wh |
power output |
24V/80mA, Output overcurrent and short circuit protection |
Current measurement, current limiting protection, voltage limiting |
<36V |
Battery power supply time |
≥ 6 hours |
Overall dimensions |
270*210*100(mm) |
total machine weight |
About 2.5 (kg) |
Pressure module interface |
7-pin aviation socket |
Maximum output/measurement rate |
20 o'clock/second |
port |
4mm standard socket, output measurement, layout of strong and weak electrical terminals in different areas, and colored identification of strong electrical terminals |
Operating height |
Below an average sea level of 3000 meters |
interface |
USB |
communication |
5G、WIFI |
data store |
8GTF card |