Welcome Customer !

Membership

Help

Zhejiang Jinyi Instrument Co., Ltd
Custom manufacturer

Main Products:

foodequippro>Products

JY228 iCCM Instrument Control Test Intelligent Calibration Instrument

NegotiableUpdate on 11/13
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Product Overview: A specialized tool for multifunctional signal detection, which realizes intelligent and paperless calibration of various cards and instruments, as well as data exchange in iCCM platform. It realizes functions such as temperature probe insulation continuity inspection, board calibration, threshold calibration, etc. Calibration equipment: Main features:

Product Details

Product Details

Feature Introduction

·support andJY7000The digital pressure module is used in conjunction to complete pressure calibration and precision measurement;

·Support simultaneous simulation and measurement of thermocouples, provide three cold end temperature compensation methods, and use an external Pt100 temperature probe for real-time temperature compensation; Using fixed temperature values for constant temperature compensation; Using built-in temperature sensors for cold end temperature compensation;

·When simulating and measuring thermocouples and thermal resistors, display the corresponding mV/Ω electrical value;

·Conversion: Supports output/measurement conversion to other data, supports linear, square, and square root conversion;

·Measurement record: Real time recording of measurement values and the ability to generate curve graphs;

·System settings: Supports network connections, and the system provides multiple software tools such as thermal engineering manuals, calculators, and unit conversions;

·Battery management: supports low battery prompt, charging prompt, battery status information display, etc., can be used while charging;

·7-inch large color industrial touch screen, no button design, can be directly operated by touch screen;

·Fashionable and high-end stress relieving straps that are both convenient to carry and free up hands during use;

·The range will be displayed on the screen, and there is no need to consult the user manual;

·Quick output: Save frequently used process calibration points, and each channel and signal can be output by pressing shortcut keys (long press to edit);

·The invention patent of "Numerical Control Variable Resistance Device" enables the calibrator to respond faster, more stably, and allow a wider excitation current when outputting resistance signals. Supports fast pulse RTD temperature transmitters andPLCCalibration;

·When outputting mA signals, two power supply methods are provided: internal and external;

·The frequency output supports both square waves and sine waves. The duty cycle and amplitude can be set for square waves, while the amplitude can be set for sine waves;

·When measuring mA signals, it can provide loop power to the transmitter;

·External 24V DC power supply, capable of supplying power to other devices;

·Switching of pressure and temperature units, mutual calculation of voltage, resistance, and temperature values, in compliance with the ITS-90 international temperature scale;

·Step: can set step level output signal;

·Slope: can set slope output signal, automatically calibrate switch instruments;

·Output and measurement have data conversion functions, and each channel and signal are independently designed;

·When measuring resistance and thermal resistance, support two -, three -, and four wire wiring methods;

·Equipped with conventional multimeter functions, it can measure signals such as frequency, diodes, switches, AC/DC voltage, capacitance, etc;

·Equipped with professional RCP63 measurement function, it can perform manual or automatic detection and calculation of thermocouple performance;

·Equipped with 50V or100VThe insulation resistance measurement function.




technical indicators

Process output (one year, ambient temperature)20℃±5℃)

Output range

accuracy

resolution

internal resistance

noise

20 V

0.01% reading+0.005% full degree

1 mV

≤0.2Ω

≤20uV

10 V

0.008% reading+0.001% full degree

100 uV

1 V

10 uV

100 mV

0.008% reading+5uV

1 uV

≤50mΩ

≤2uV

20 mA

0.01% reading+1uA

1 uA

≥20MΩ

≤0.1uA

4000 Ω

0.008% reading+0.001% full degree

100 mΩ

External excitation current: 0.1mA

400 Ω

10

External excitation current: 1mA

50000 Hz

0.004% full degree

1 Hz

Square wave and sine wave;
Square wave duty cycle adjustment range: 5% to 95%

5000 Hz

0.1 Hz

500 Hz

0.01 Hz

Process measurement (one year, ambient temperature)20℃±5℃)

Measurement range

accuracy

resolution

internal resistance

10 V

0.008% reading+0.001% full degree

0.1 mV

1 MΩ

20 mA

0.01% reading+1uA

1 uA

10 Ω


Temperature probe measurement (one year, ambient temperature)20℃±5℃)

Measurement range

accuracy

resolution

internal resistance

±100 mV

0.008% reading+5uV

1 uV

≥500 MΩ

500 Ω

0.008% reading+0.001% full degree

10 mΩ

/

5000 Ω

100 mΩ

/

Insulation testing (one year, ambient temperature)20℃±5℃)

output voltage

display range

resolution

accuracy

50 V/100 V

0~9.99 MΩ

0.01 M

± (10% reading+5 words)

10.0~99.9 MΩ

0.1 M

100~200 MΩ

1 M

Multimeter module (one year, ambient temperature)20℃±5℃)

Measurement range

range

accuracy

capacitance

2.000 uF

± (4% reading+50 words)

20.00 uF

200.0 uF

AC voltage

5.000 V

± (1% reading+50 words)

50.00 V

500.0 V

DC voltage

5.000 V

± (0.1% reading+5 words)

50.00 V

500.0 V

diode

/

discrete signal

On/off or level discrimination

/

frequency

50000 Hz

0.004% full degree

9999.9 Hz

999.99 Hz

RTD thermistor (one year, ambient temperature 20 ℃± 5 ℃)

Output/Measurement

temperature range

accuracy

Pt100

(-200~850)℃

0.12 ℃

Pt1000

(-200~850)℃

0.12 ℃

Cu50

(-40~150)℃

0.05 ℃

G

(-40~150)℃

0.05 ℃

BA1

(-190~650)℃

0.12 ℃

BA3

(-40~100)℃

0.06 ℃


TC thermocouple (one year, ambient temperature: 20 ° C)±5

Output/Measurement

range

accuracy

S

(-50~0)℃

1.27℃

(0~100)℃

0.93℃

(100~1768)℃

0.68℃

R

(-50~0)℃

1.36℃

(0~200)℃

0.95℃

(200~1768)℃

0.58℃

B

(600~800)℃

0.92℃

(800~1820)℃

0.69℃

K

(-250~-200)℃

1.14℃

(-200~-100)℃

0.36℃

(-100~600)℃

0.18℃

(600~1372)℃

0.28℃

N

(-200~-100)℃

0.54℃

(-100~1300)℃

0.25℃

E

(-250~-200)℃

0.60℃

(-200~-100)℃

0.23℃

(-100~700)℃

0.12℃

(700~1000)℃

0.15℃

J

(-200~-100)℃

0.30℃

(-100~700)℃

0.13℃

(700~1200)℃

0.19℃

T

(-250~-100)℃

0.87℃

(-100~0)℃

0.19℃

(0~400)℃

0.13℃

WRe3-25

(0-2315)℃

0.91℃

WRe3-26

(0-2315)℃

0.89℃



Other indicators

project

Specifications

Process simulation quantity temperature influence quantity

<(10ppm reading+5ppm quantity)
Cheng/℃

Verification table working environment

0-45℃/0-90%rh

Calibration table storage temperature

-20~55℃

Voltage output carrying capacity

>10mA, Output overcurrent and short circuit protection

power supply

DC 24V/1.6A

battery

14.4V 6000mAh/86.4Wh

power output

24V/80mA, Output overcurrent and short circuit protection

Current measurement, current limiting protection, voltage limiting

<36V

Battery power supply time

≥ 6 hours

Overall dimensions

270*210*100(mm)

total machine weight

About 2.5 (kg)

Pressure module interface

7-pin aviation socket

Maximum output/measurement rate

20 o'clock/second

port

4mm standard socket, output measurement, layout of strong and weak electrical terminals in different areas, and colored identification of strong electrical terminals

Operating height

Below an average sea level of 3000 meters

interface

USB

communication

5G、WIFI

data store

8GTF card