Feature Overview
It can program commonly used step signals and step interval time, and supports two modes: automatic step or manual step;
Multi functional and programmable unit conversion function, which can convert V, mA output or measurement into other unit quantities;
When outputting mA signals, it supports two modes: analog transmitter and mA source;
Provide 24V DC circuit power supply and measure the mA signal in the circuit;
When outputting Hz signals, the amplitude can be set and supports two modes: square wave or sine wave;
Support signal fine-tuning on any digit of the output value, used for calibration of pointer instruments;
Can lock or unlock measurement values during measurement;
When measuring values, collect the minimum and maximum values and calculate the average.
Accurate and stable
Small temperature drift: ± (0.0002% reading+0.0002% range)/℃;
Small time drift: ± (0.004% reading+0.001% range)/year;
Low voltage noise: 1 × 10-5Range;
Low current noise: 1 × 10-5Range;
Strong resistance to various interferences;
Original 24K gold-plated low thermoelectric potential material, with more stable signal.
safe and reliable
Both the output and measurement ports can withstand 250V AC ingress;
Both the output port and 24V DC port can automatically recover from short circuits;
The measurement channels are all automatic range;
The charging system has automatic safety protection such as overcharging, overdischarging, and short circuit;
The shell has high insulation strength and strong seismic and drop resistance.
Key Metrics
Output and measurement (one year, ambient temperature 20 ℃± 5 ℃) ★ represents the basic range range range
| Output range | Maximum allowable error of 0.01 level | Maximum allowable error of 0.02 level | Maximum allowable error at level 0.05 | resolution |
| 10.99999 V | ± (0.008% reading+0.0001 V) | ± (0.015% reading+0.0001 V) | ± (0.035% reading+0.0001 V) | 10 μV |
| 1.099999 V ★ | ± (0.008% reading+0.00001 V) | ± (0.015% reading+0.00001 V) | ± (0.035% reading+0.00001 V) | 1 μV |
|
– 99.9999 mV ~109.9999 mV |
± (0.008% reading+0.003 mV) | ± (0.015% reading+0.003 mV) | ± (0.035% reading+0.003 mV) | 1 μV |
| 30.0999 mA ★ | ± (0.01% reading+1 μ A) | ± (0.015% reading+1 μ A) | ± (0.035% reading+1 μ A) | 0.1 μA |
| 54999.9 Hz ★ | ± 2 Hz | ± 2 Hz | ± 2 Hz | 0.1 Hz |
| 5499.99 Hz | ± 0.2 Hz | ± 0.2 Hz | ± 0.2 Hz | 0.01 Hz |
| 549.999 Hz | ± 0.02 Hz | ± 0.02 Hz | ± 0.02 Hz | 0.001 Hz |
| 24V DC circuit power supply | ± 5% | ± 5% | ± 5% | |
| Maximum output current of V: 10mA; 20mA output load capacity ≤ 1k Ω; 50000Hz load capacity ≥ 10k Ω; Maximum output current of 24V DC circuit power supply: 80mA. | ||||
| Measurement range | Maximum allowable error of 0.01 level | Maximum allowable error of 0.02 level | Maximum allowable error at level 0.05 | resolution |
| ± 59.9999 V | ± (0.008% reading+0.0005 V) | ± (0.015% reading+0.0005 V) | ± (0.035% reading+0.0005 V) | 0.1 mV |
| ± 5.99999 V ★ | ± (0.008% reading+0.00005 V) | ± (0.015% reading+0.00005 V) | ± (0.035% reading+0.00005 V) | 0.01 mV |
| ± 599.999 mV | ± (0.008% reading+0.005 mV) | ± (0.015% reading+0.005 mV) | ± (0.035% reading+0.005 mV) | 1 μV |
| ± 119.999 mV | ± (0.008% reading+0.003 mV) | ± (0.015% reading+0.003 mV) | ± (0.035% reading+0.003 mV) | 1 μV |
| ± 119.999 mA | ± (0.01% reading+1 μ A) | ± (0.015% reading+1 μ A) | ± (0.035% reading+1 μ A) | 1 μA |
| ± 23.9999 mA ★ | ± (0.01% reading+1 μ A) | ± (0.015% reading+1 μ A) | ± (0.035% reading+1 μ A) | 0.1 μA |
| 59999.9 Hz ★ | ± 2 Hz | ± 2 Hz | ± 2 Hz | 0.1 Hz |
| 9999.99 Hz | ± 0.4 Hz | ± 0.4 Hz | ± 0.4 Hz | 0.01 Hz |
| 999.999 Hz | ± 0.04 Hz | ± 0.04 Hz | ± 0.04 Hz | 0.001 Hz |
| 50V measurement input impedance: ≥ 1M Ω; 5V measurement input impedance ≥ 500M Ω; current measurement input impedance ≤ 10 Ω; Hz measurement input impedance ≥ 1M Ω 1Hz~100Hz:0.2V~250V(rms); 100Hz~10kHz:0.25V~30V(rms); 10kHz~59kHz:0.5V~30V(rms) | ||||
Other indicators
| project | Specifications | |
| warm-up time | After 5 minutes of startup, the technical specifications are met | |
| Electrical measurement connection | 4mm gold-plated lantern plug | |
| Side port connection | Pressure/24V DC HART/RS232 connection |
aviation plug |
| Charger connection | φ (5.5 × 2.5) mm DC port | |
| Overall dimensions | (205x110x52)mm | |
| weight | About 0.9kg | |
| display screen | 3.7-inch monochrome dot matrix LCD display screen | |
| Display Contrast | adjustable | |
| button backlight | LED, closable | |
| Battery Category Capacity | Lithium battery pack 7.4V/4800mAh, 36Wh | |
| Battery lifespan | Charge and discharge more than 300 times | |
| Battery charging time | About 4 hours | |
| Battery usage time | 12 hours (measurement and output, turn on circuit power 12mA, screen and button backlight on); >15 hours (measurement and output, screen and button backlight on); 16 hours (measurement or output, screen and button backlight on, continuous). |
|
| Adapter input | (100-240)V AC,(50-60)Hz | |
| Adapter output | 8.6V DC/1A | |
| Operating Temperature | (-10~50)°C | |
| Storage temperature | (-20~60)°C | |
| Maximum relative humidity without condensation | 90% at 35 ° C; At 75% and 40 ° C; 45% at 50 ° C. |
|
Optional features
Secondary instrument calibration software
-National metrological verification regulations with multiple instruments, unified management of relevant information of the tested instruments;
-Automatically execute the calibration process, analyze data, and generate calibration certificates (calibration reports) and other reports;
-Can import custom report templates (Excel documents) and manage reports and templates uniformly according to customer verification requirements;
-The on-site detection data stored in the calibrator can be uniformly managed for data copying, editing, storage, printing, etc;
-The calibrator can be operated remotely through a computer, and can perform signal source output or signal acquisition work for a long time, while managing corresponding data;
-Provides signal conversion tools for 10 types of thermocouples and 8 types of thermal resistors.