Condumax II is a fully integrated hydrocarbon dew point analyzer that utilizes mature black spot measurement technology that has been successfully validated on the first generation Condumax products
| |
Condumax II is a fully integrated hydrocarbon dew point analyzer that utilizes mature black spot measurement technology that has been successfully validated on the first generation Condumax products. In the sturdy and durable explosion-proof casing, not only is there a hydrocarbon dew point measurement unit, but also a water dew point module can be combined simultaneously. This product only requires an external main power supply to be put into operation. Condumax II provides on-site continuous measurement reading display, as well as analog and digital outputs for remote monitoring.
highlight
- Fully automatic online analysis
- Objective measurement with good repeatability
- Basic Cold Mirror Principle
- Patent testing technology
- self-cleaning
- CSA certification
- ATEX authorized certification Ex II 2 G EExd IIB+H2 T4 can be used in hazardous environments
- Sampling system/sensing unit design suitable for outdoor on-site installation
- No need to blow the carrier gas or external refrigeration device
- Optional water dew point integration module for hydrocarbon dew point analyzer
- Modbus RTU transmission protocol, touch screen interface
|
 |
technical indicators
|
| sensor |
| measurement technology |
Black Spot DARK SPOT ™ Fixed sample gas analysis. Direct optical path detection of hydrocarbon condensation at the dew point temperature point of hydrocarbons |
| Sensor cooling |
Automatic cooling through 3-level Peltier semiconductor technology |
| maximum range |
-34 ° C HCdp at an ambient temperature of+21 ° C |
| precision |
± 0.5 ° C hydrocarbon dew point |
| Sample gas flow rate |
0.03m³/hr (0.5 Nl/min) |
| Dew point detection (optional) |
| precision |
± 1 ° C-59~+20 ° C dew point ± 2 ° C -100~-60 ° C dew point |
| Sample gas flow rate |
1 ~ 5 Nl/min |
| pressure measurement |
| model |
Hydrocarbon dew point channel PDCR 1050-3465: 0~100 barg Water dew point channel PDCR 1050-3574: 0~210 barg |
| precision |
±0.25% FS |
| Hydrocarbon dew point analyzer |
| gas source |
Natural gas, with a maximum pressure of 100 barg, the sampling system includes a pressure regulator valve |
| shell |
EExd explosion-proof enclosure with detachable transparent window panel. Internal insulation to prevent condensation formation |
| Sample gas interface |
¼' NPT (f) |
| working environment |
Indoor/Outdoor -20~60 ° C. Maximum 95% RH |
| power supply |
90-264 V AC 50/60 Hz, Host 200 W; Indoor sampling system 300 W; Outdoor sampling system 400 W |
| weight |
Host 22.5kg Sampling system (indoor) 42kg (approximately) Sampling system (outdoor) 57kg (approximately) |
| Integrated display |
touch screen display |
| Output and alarm |
Modbus RTU, RS485 @ 9600 baud rate Two 4-20 mA linear (non isolated) outputs, allowing users to freely choose dew point or pressure parameters. Hydrocarbon and water dew point alarms are triggered through software registers. Each channel can integrate a low flow alarm module. |
Options and accessories:
The remote interface of Condumax II integrates a control module for security scenarios, providing display functions for all parameters, chart and data acquisition functions, parameter control, and remote diagnostic functions.
19 'standard rack installation with 6.5' color LCD display and mouse operation.
Dual channel remote digital display, providing readings of selected parameters (any two of hydrocarbon dew point, water dew point, or pressure readings), high brightness and 3.5-bit LED display, with secondary analog output and dual channel alarm.
The factory level control integration ActiveX module integrates software programs into a universal factory level software system through the Modbus protocol. |
|
|
|
|
|
|