WBE-QUV-3 UV Weathering Resistance Tester

NegotiableUpdate on 10/03
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

This testing machine can simulate the hazards caused by sunlight, rain, and condensation. It uses fluorescent ultraviolet lamps to simulate sunlight exposure and sprays tap water or distilled water to simulate the effect of rain. The tested material is placed in a light cycle program at a certain temperature for testing. This machine can test the hazards that occur outdoors for months or years in a matter of days or weeks. The types of hazards include fading, discoloration, loss of light, powdering, cracking, turbidity, bubbles, brittleness, reduced strength, peeling, decay, and oxidation.

Product Details

Product Features

Compliant with standards:

GB/T 14522-93 National Standard of the People's Republic of China "Accelerated Test Methods for Plastics, Coatings, Rubber Materials, and Artificial Climate for Mechanical Industry Products"

● ASTM G 153 ● ASTM G 154 ● ASTM D 4329 ● ASTM D 4799 ● ASTM D 4587 ● SAE J 2020 ● ISO 4892

Technical Specifications

model

WBE-QUV-3

Studio dimensions: width, depth, height

W1130mm×D500mm×H400mm

Dimensions: Width, Depth, Height

W1300mm×D700mm×H1400mm

Test function

Light/rain/condensation/irradiation

Irradiation lamp tube

A-340 or B-313 8 horizontal emissions in a row

Test sample capability

1140 × 500 × 400mm testing space does not require sampling

Lamp brand

Original imported ATALS from the United States

Center distance of the lamp


The closest parallel distance between the test piece and the lamp surface

50-400mm adjustable

wavelength range

UV-A(315~400nm)、UV-B(280~315nm)

radiation intensity

Adjustable within 0.3W/m2 to 1W/m2 (optional)

Temperature resolution

0.1℃

Light temperature range

50 ℃~70 ℃/Temperature tolerance is ± 3 ℃

Condensation temperature range

40 ℃~60 ℃/Temperature tolerance is ± 3 ℃

temperature control

PID self-tuning temperature control method

humidity range

About 45%~70% R.H (illuminated state)/98% or more (condensed state)

Sink requirements

The water depth is not greater than 25mm and there is an automatic water supply controller

Standard specimen size/testing space

1140×500×400mm

Suggested instrument usage environment

5~35 ℃, 40%~85% R · H, 300mm away from the wall