Detection Principle
The G22 type total phosphorus rapid analyzer is developed and designed based on the "Determination of Total Phosphorus in Water Quality - Ammonium molybdate spectrophotometric method". It uses a closed digestion tube and potassium persulfate as an oxidant to convert all phosphorus containing compounds in the sample into phosphate ions at 125 ℃. Then, under acidic conditions, orthophosphate reacts with ammonium molybdate and potassium antimony tartrate to generate phosphomolybdic acid, which is reduced by the reducing agent ascorbic acid to form a blue complex. The total phosphorus detection item is selected by colorimetric method. The imported cold light source emits a light source signal to the sensor, and the data is processed by RAM processor for direct feedback of the sample concentration value, which is displayed in mg/L.
Product Features
01. Implementation standard: Implement the national standard GB11893-89 to ensure accurate and effective measurement.
02. Accurate detection: Developed according to national testing standards, widely applicable for the detection of production and domestic wastewater.
03. Advantages of light source: Imported cold light source is used, with excellent optical performance and a lifespan of up to 100000 hours.
04. Intelligent operation: The device adopts guided operation, and users can easily complete detection operations according to prompts.
05. Intelligent printing: It can be set to automatically print density results upon completion of detection, or select manual printing mode.
06. Continuous measurement: With multiple samples to be tested, the continuous detection mode can be activated to achieve higher detection efficiency.
07. High cost-effectiveness: Equipped with complete consumables and reagents, low price and minimal usage.
08. Pre made reagents: To provide customers with pre made reagents, simply add the sample to detect the value.
09. Operational safety: Dissolve and compare colors in one, without the need to replace the colorimetric dish for detection, the measurement method is simple and efficient.
10. Intelligent program: 4.3-inch color LCD screen display, combined with the UI interaction system developed by Greencairui, simple and time-saving operation.
Total phosphorus rapid analyzer (water quality analysis)
Technical Parameter
| PRODUCT MODEL | G22 type total phosphorus rapid analyzer |
| Detection parameters | total phosphorus |
| Measurement standards | GB11893-89 Determination of total phosphorus in water quality Ammonium molybdate spectrophotometric method |
| measuring range | 0-16mg/L |
| limit of detection | 0.02mg/L |
| Dissolving temperature | 150 ℃, 15 minutes |
| detection time | 20 minutes/10 samples |
| Indication error | ≤ 5% or ± 0.04mg/L |
| Instrument power supply | AC(220V±5%),50Hz |
| work environment | 5-40 ℃, ≤ 85% non condensing |
| Instrument dimensions | 300mm×230mm×115mm |
| Instrument weight | About 2.8kg |
| Display mode | 4.3-inch high-definition color LCD screen |
| Colorimetric method | Colorimetric tube (digestion colorimetric integrated) |
| Expand functionality | PC online |