Welcome Customer !

Membership

Help

Beijing Tian'an United Technology Co., Ltd
Custom manufacturer

Main Products:

foodequippro>Products

Beijing Tian'an United Technology Co., Ltd

  • E-mail

    1347609469@qq.com

  • Phone

    15201117305,15321330700

  • Address

    No. 13 Linyin North Street, Pinggu District, Beijing

Contact Now

Non metallic digital ultrasonic flaw detector

NegotiableUpdate on 02/02
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The digital ultrasonic flaw detector measures various internal defects (cracks, inclusions, pores, etc.) of non-metallic materials (refractory bricks, graphite electrodes, plastics, ceramics, rubber, plywood, etc.) quickly, conveniently, without damage, and accurately.

Product Details

TA9106Digital non-metallic flaw detector

I. Overview

Our company has newly developed a cost-effective productNon metallic digital ultrasonic flaw detectorIt effectively solves the problem of utilizationUltrasonic flaw detectorMeasure each

Non metallic species(Refractory bricks, graphite electrodes, plastics, ceramics, rubberplywoodwait)Internal defects in the material(Cracks, inclusions, pores, etc)problemoptional

Hollow or solid waveforms allow operators to see more clearly. This machine can quickly, conveniently, without damage, and accurately detect various non-metallic internal defects.

2、 Main functions

5.7Inch true colorIPSLCD screen display (wide viewing angle),640x480High definition display with high resolution;

There are two measurement methods: continuous measurement method(Free running mode)And single measurement method;

It has three measurement modes: high-precision measurement mode can take the average of multiple measurements and automatically filter suspicious data to ensure more accurate and stable measurement values; The fast measurement mode can achieve real-time scanning function;

Equipped with temperature compensation function: Real time temperature compensation technology can automatically compensate for measurement errors caused by changes in ambient temperature and probe temperature, making measurements more accurate;

There are five statistical measures: mean(MEAN), maximum value(MAX)Minimum value(MIN)Test frequency(NO.)Standard deviation(S.DEV);

Zero point calibration, single point calibration, or two-point calibration methods can be used to calibrate the instrument,And basic calibration and temperature coefficient calibration methods can be used to calibrate the system of the measuring head

Correction of statistical errors;

Has storage function: can store up to500A measurement value;

It has a deletion function: it can delete individual suspicious data that appears during measurement, as well as all data in the storage area, in order to conduct new measurements;

Can set limit: automatic alarm for measurement values outside the limit;

Equipped with power indicator function;

There is a buzzing sound during the operation process;

There are three shutdown modes available: manual shutdown mode, timeout automatic shutdown mode, and low battery automatic shutdown mode, and the timeout automatic shutdown waiting time can be set;

3、 Technical parameters

name

Technical data

Time range of sound

15~9999MS

Pulse shift(ms

Pulse shift(ms):1+9999 Adjust step distance:1ms

Launch power supply voltage

negative500V

Transmitting pulse width

Negative pulse square wave with adjustable pulse width100NS-9990NS

Transmitting repetition frequency

100/300Hz

Probe damping resistor

about1KΩ

Digital sampling

8bitreal-time100MHz

display mode

radio frequency

display device

TFTIndustrial color LCD display screen(640x4805.7 Inches),LEDbacklight

LCD screen refresh rate

60Hz

Equipped with probe

50KP28FSpecialized probe 1 right

working method

Dual probe (one receiver and one transmitter)

Frequency range(MHz

40KHz ~ 1MHz

Gain adjustment(dB

0120 Adjust step distance:0.21246810,20,50

Linear inhibition

Screen height080%Step distance:1%

Vertical linear error

Vertical linear error not greater than3%

Horizontal linear error

Within the scanning range, not greater than0.2%

call the police

Incoming wave alarm Wave loss alarm

A-Scandisplay area

Full screen or partial A-ScanDisplay freeze and thawA-Scanfill

Testing channel

300a

data storage

500groupA-Scangraphics

unit of measurement

Mm

4、 Meet the standards

1. JJG 746-2004National Metrological Verification Regulations of the People's Republic of China-Ultrasonic flaw detector

2. JB/T 10061-1999AGeneral technical specifications for pulse reflection ultrasonic flaw detector

3. JB/T 10062-1999Method for Testing the Performance of Ultrasonic Testing Probes

4. JB/T 9214-1999AMethod for Testing the Performance of Pulse Reflection Ultrasonic Testing System

5. Z2344-93Method for Pulse Reflection Ultrasonic Testing of Metallic Materials

● European standards(EN12668)It includes three parts:

1. EN12668-1Non-destructive testing-Characteristics and certification of ultrasonic testing equipment-1Part: Instruments

2. EN12668-2Non-destructive testing-Characteristics and certification of ultrasonic testing equipment-2Part: Probe

3. EN12668-3Non-destructive testing-Characteristics and certification of ultrasonic testing equipment-3Part: Comprehensive equipment