ND2010A typeFully automatic pour point analyzerThe system adopts modern high-tech microelectronic control technology, uses MCS-51 series single chip microcomputer as the system control core, and is equipped with a 6-inch 256 color intelligent color screen display, perpetual calendar clock timing display and screen type touch keys to ensure the progressiveness and reliability of the whole machine; Chinese LCD display, human-machine menu dialogue, guided operation,Measurement process * according toGB/510 "Determination of Condensation Point of Petroleum Products"andGB/3535 "Petroleum pour point determination method"The method requires measurement, and the measurement process is fully automated
ND2010A typeFully automatic pour point analyzerWhen the temperature of the sample reaches the detection temperature, the test tube containing the sample is automatically tilted at 45 ℃. Infrared fiber optic cable sensing technology is used to detect the flowability freezing point of the sample, cool and control the temperature, and automatically print records to achieve automated analysis of the sample.
ND2010A typeMain technical indicators
⑴. Scope of application: Petroleum products with pour point ranging from 30 ℃ to -70 ℃
⑵. Measurement repeatability: pour point 2 ℃, pour point 3 ℃
⑶. Display resolution: 0.1 ℃ Accuracy: 0.3%
⑷. Refrigeration limit depth: below -65 ℃ (double compressor cascade refrigeration)
⑸. Cooling speed: 10 minutes temperature difference>15 ℃
⑹. Sample dosage: 30mL per time
⑺. Sample quantity: 1
Rated power: 1000W
9. Power supply voltage: 220V ± 10% 50Hz ± 1Hz
(10) Operating environment: temperature 25 ℃± 10 ℃; Relative humidity<85% PH
11. Dimensions: 400 × 750 × 600mm (width × height × depth)
⑿. Weight: 35Kg
ND2010A typeWorking principle
Microcomputer intelligent control has achieved full automation of the measurement process, that is, automatically heating the sample to 50 ℃, naturally cooling it to 35 ℃, and then automatically placing the sample into a cold trap. When the sample temperature reaches the detection temperature, the cold bath box containing the sample is automatically tilted at 45 ℃, and the flowability freezing point of the sample is detected using red outer optical fiber cable sensing technology. The entire process of detecting the sample's solidification is * every time; This ensures that the solidification point of the sample is accurate. The program control realizes the measurement of the solidification point of the sample (i.e. the sample flows when the temperature is 2 ℃ higher than this temperature, and solidifies when the temperature is equal to this temperature); And automatically store and print records of data.