E-mail
2095778920@qq.com
-
Phone
15506020928
-
Address
No. 3 Tongle Hutong, Yueshan Road, Jilin City, Jilin Province
Jilin Jifen Instrument Co., Ltd
2095778920@qq.com
15506020928
No. 3 Tongle Hutong, Yueshan Road, Jilin City, Jilin Province
| brand | other brands | price range | 5000-10000 |
|---|---|---|---|
| Origin category | domestic | application field | petroleum |
KV-500A typePetroleum product kinematic viscosity testerCompliant with the national standard GB/T265-88 "Determination of Kinematic Viscosity of Petroleum Products and Algorithm for Dynamic Viscometers", it is used to determine the kinematic viscosity and dynamic viscosity of liquid petroleum products (referring to Newtonian liquids) (dynamic viscosity can be obtained by multiplying the measured kinematic viscosity by the density of the liquid at the same temperature), with the unit of m/s2, usually used in practice in mm/s2. This instrument is also applicable to the methods specified in ISO3104, ASTM D445, and IP71 standards.Suitable for determining the kinematic viscosity of liquid petroleum products.
KV-500A typePetroleum product kinematic viscosity testerMain technical indicators:
⑴. Capillary Viscometer: One set (13 pieces), all in accordance with SH/T0173-92 "Technical Conditions for Glass Capillary Viscometers", with inner diameters of 0 4. 0 6. 0 8. 1 0, 1 2. 1 5. 2 0, 2 5. 3 0, 3 5. 4 0, 5 0, 6 0 millimeters (including 5.0 and 6.0 without coefficient table, users can calibrate by themselves)
⑵. Temperature control points: 20, 40, 50, 80, 100 ℃ Temperature control accuracy: ± 0 1 ℃ digital display temperature, resolution: 0 Accuracy at 1 ℃: ± 0 1%
⑶. Temperature controlled heater power: 600W
⑷. Auxiliary heater power: 800W
⑸. Mixing motor power: 25W
Mixing speed: 1400 revolutions per minute
⑺. Number of measurement holes: 4
⑻. Timing time: 999.9S
Timing accuracy: 0 1S
10. Power supply voltage: AC220V ± 10% 50Hz ± 1Hz
11. Operating environment: temperature 25 ℃± 10 ℃; Relative humidity<85%
KV-500A typeThe working principle isNewtonian liquids refer to liquids where the ratio of shear stress to shear rate is a constant, and also where the shear stress and shear rate of a liquid are independent of viscosity.The instrument is based on the MCS-51 series microcontroller. After setting a constant temperature point, the instrument automatically heats up, maintains a constant temperature, and turns on the stirring motor. When the temperature is constant, the test can be carried out according to the method specified in GB/T265-88 standard.