Gasoline car engine oil oxidation stability testerSuitable for evaluating the oxidation stability of new and in use turbine oil with the same composition (base oil and additives). It can also be used to evaluate new mineral insulating oils containing 2,6-di-tert-butyl-p-cresol as a rapid method for assessing their oxidation stability. The test results are used to verify the performance of new mineral insulating oils containing 2,6-di-tert-butyl-p-cresol or 2,6-di-tert-butylphenol or both, and can be continuously controlled for each batch of oil. Mainly used in the fields of petroleum, chemical, power, pharmaceuticals, commercial inspection, scientific research, environmental protection, etc.
Gasoline car engine oil oxidation stability tester
Adaptation standards: SH/T0193, ASTM D 2272, D4742, IP229
Instrument Features
1. Used for new oil, in use oil, steam turbine oil, lubricating oil, insulation oil, and automotive gasoline engine oil, etc.
2. The metal measuring cell reduces the unpleasant odor emitted by the medium oil during testing, avoiding potential dangers that may arise due to prolonged experimental time.
3. Independent operation design enables the testing process to have the ability to quickly turnover, greatly improving work efficiency when using multiple devices.
4. Desktop design has a small volume, and the metal bath does not require the use of liquid media.
5. Easy ventilation, no need to place in a fume hood, exhaust gas and odors can be easily discharged through plastic pipes.
6. Sample temperature sensor component: It can measure the temperature of the sample during the testing process.
Technical Specifications
• Power supply: AC220V ± 10%
Heating tube power: 2500W
Pressure sensor range: 0-1.6 MPa, accuracy: ± 2 ‰
• Temperature control points: 140.0 ℃, 150.0 ℃
Temperature control accuracy: ± 0.1 ℃ Environmental temperature: 5 ℃~40 ℃
Rotating mechanism speed: 100 ± 5r/min
• Angle between oxygen bomb and horizontal plane: 30 ° •
• Dimensions: 760mm × 390mm × 310mm
• Instrument weight: 35kg