E-mail
359845197@qq.com
-
Phone
13918091972
-
Address
508, Building 4, Lane 680, Shuishe West Road, Baoshan District, Shanghai
Shanghai Wangxu Electric Co., Ltd
359845197@qq.com
13918091972
508, Building 4, Lane 680, Shuishe West Road, Baoshan District, Shanghai
HK-3013SB Closed Flash Point Tester
The whole machine adopts an advanced ATMEL AT91SAM9G10 microprocessor as the core operating system, with a large touch screen and a human-machine dialogue interface displayed in both Chinese and English. It can pre value parameters such as flash point temperature, sample number, atmospheric pressure, and test date. It has a prompt menu and guided input, which is convenient and fast. With advanced technology and high technological content, it is an ideal substitute for imported products, and the product is at the domestic level. Used to determine the closed flash point value of petroleum products, it is widely used in the railway, aviation, power, petroleum industries, and scientific research departments.
characteristic
Adopting a 7-inch 640 × 480 true color touch screen, displaying in both Chinese and English, with a help menu for easy operation.
Simulate and track the function curve of temperature rise and test time, with software prompts for modification in case of misoperation
The temperature exceeds the value and automatically stops heating. After the test, the gas source will be automatically cut off
The electronic ignition device adopts a new type of silicon nitride igniter, which is closer to an open flame, reliable in repeatability, and has a long service life.
Gas ignition with automatic ignition, no need for manual ignition
The instrument has self checking, diagnostic, fault alarm and security functions;
Built in atmospheric pressure sensor, automatic pressure correction, no need to manually set atmospheric pressure value.
Equipped with RJ45 interface for LAN data transmission and firmware upgrade
Equipped with USB interface for software upgrade and data export
Can store 1000 test records, can be viewed and printed by operation date
HK-3013SBClosed flash point tester
Relative humidity of the environment: ≤ 85%
Power supply: AC220V ± 10%, frequency: (50-60) Hz;
There are vibrations around that do not affect performance;
There is no corrosive gas present in the surrounding air;
There are no other magnetic or electric fields around that can cause influence, except for the geomagnetic field.
Under the conditions specified in the national standard GB/T261-2008, the instrument places the sample into a test cup and heats the test cup containing the test oil. The lowest temperature at which the condensed synthesis gas formed by the petroleum vapor and the surrounding air ignites upon contact with the flame is used as the flash point.
The computer sends instructions through the I/O port based on the collected temperature changes to control the heater, causing the temperature of the test oil to rise at a certain rate. The mixing speed, lid opening ignition, ignition time, flash point detection, etc. are all automatically controlled. When flash fire is detected, the computer system stops data collection, displays the flash fire temperature, prints the recorded results, stops heating, turns off the flame, and the boom automatically lifts, ending the test.
1. Working power supply: AC 220V ± 10%, (50-60) Hz
2. Whole machine power consumption: ≤ 300W
3. Range: Room temperature to 370 ℃;
4. Accuracy: ≤ 110 ℃± 1 ℃
≥110℃±2℃
5. Repeatability: ≤ 2 ℃
6. Resolution: 0.1 ℃
7. Ignition method: electronic ignition, automatic gas flame.
8. Cooling method: forced air cooling
9. Dimensions and weight: Control host: L270mm × W170mm × H130mm, approximately 2.0kg
Measurement section: L480mm × W245mm × H280mm, approximately 18.0kg
Control host: 1 unit
Measurement section: 1 unit
Closed oil cup: 1
Igniter: 1 piece
Power cord: 1 piece
Power adapter cable: 1 piece
Data connection cable: 1 piece
Printing paper: 1 roll
Instruction manual: 1 copy
Certificate of conformity: 1 copy
Warranty card: 1 copy