COD ammonia nitrogen total phosphorus analyzer (water quality analysis)
Detection Principle
G63 COD ammonia nitrogen total phosphorus analyzer, conventional detection item detection principle: COD is determined by closed catalytic digestion colorimetric method using a digestion tube, ammonia nitrogen is determined by Nessler's reagent colorimetric method, total phosphorus is determined by ammonium molybdate spectrophotometric method, and total nitrogen is determined by alkaline persulfate digestion spectrophotometric method. The water quality detection item is selected, and the imported cold light source emits a light source signal to the sensor. The data is processed by a RAM processor and directly fed back to the sample concentration value, which is displayed in mg/L.
COD ammonia nitrogen total phosphorus analyzer (water quality analysis)
Product Features
01. Multi parameter detection: supports the determination of multiple parameters such as COD, ammonia nitrogen, and total phosphorus.
02. Accurate detection: Developed according to national testing standards, widely applicable for the detection of production and domestic wastewater.
03. Advantages of light source: Imported cold light source is used, with excellent optical performance and a lifespan of up to 100000 hours.
04. Intelligent operation: The device adopts guided operation, and users can easily complete detection operations according to prompts.
05. Intelligent printing: It can be set to automatically print density results upon completion of detection, or select manual printing mode.
06. Continuous measurement: With multiple samples to be tested, the continuous detection mode can be activated to achieve higher detection efficiency.
07. High cost-effectiveness: Equipped with complete consumables and reagents, low price and minimal usage.
08. Pre made reagents: To provide customers with pre made reagents, simply add the sample to detect the value.
09. Operational safety: Dissolve and compare colors in one, without the need to replace the colorimetric dish for detection, the measurement method is simple and efficient.
10. Intelligent program: 4.3-inch color LCD screen display, combined with the UI interaction system developed by Greencairui, simple and time-saving operation.
Technical Parameter
| PRODUCT MODEL | G63 COD ammonia nitrogen total phosphorus analyzer | ||
| Detection parameters | COD | Ammonia nitrogen | total phosphorus |
| Measurement standards | HJ/T399-2007 | HJ535-2000 | GB11893-89 |
| measuring range | 0-15000mg/L | 0-20mg/L | 0-20mg/L |
| limit of detection | 5mg/L | 0.05mg/L | 0.02mg/L |
| Dissolving temperature | 165 ℃, 20 minutes | Not dissolving | 150 ℃, 15 minutes |
| detection time | 25 minutes/10 samples | 5 minutes/25 samples | 20 minutes/10 samples |
| Indication error | ≤ 5% or ± 4mg/L | ≤ 5% or ± 0.1mg/L | ≤ 5% or ± 0.04mg/L |
| Instrument power supply | AC(220V±5%),50Hz | ||
| work environment | 5-40 ℃, ≤ 85% non condensing | ||
| Instrument dimensions | 300mm×230mm×115mm | ||
| Instrument weight | About 2.8kg | ||
| Display mode | 4.3-inch high-definition color LCD screen | ||
| Colorimetric method | Colorimetric tube (digestion colorimetric integrated) | ||
| Expand functionality |