Phone
18321902003
-
Address
12F, Dongjin International Building, No. 501 Jinyuan 4th Road, Jiading District, Shanghai
Spectral Quality Analysis and Testing Technology (Shanghai) Co., Ltd
18321902003
12F, Dongjin International Building, No. 501 Jinyuan 4th Road, Jiading District, Shanghai
The AA3581 atomic absorption spectrophotometer has stable analysis accuracy and a user-friendly design to meet the needs of different customers. And according to different analysis needs, graphite furnace control, automatic sampler, hydride generator, etc. can be selected.
Performance Characteristics.PERFORMANCE CHARACTERISTICS
●Four or six light positions are optional: automatic 3D switching, energy balance, simultaneous preheating of multiple light positions to save waiting time, deuterium lamp clip background and stable performance element lamp can be selected.
● Automatic instrument settings: The instrument has an automatic memory function and can complete peak scanning with just one click.
Multi format saving and printing of data reports: convenient for data query and comparison.
Software design makes operation simple: The instrument software adopts a single interface multifunctional window design, real-time monitoring of data changes, supports mainstream Windows operating systems, supports remote control and calibration, and Chinese/English versions are optional.
Multiple protection system: Automatic alarm will be triggered in case of abnormal conditions such as insufficient pressure, air leakage, and engine shutdown.
Technical Specifications.TECHNICAL PARAMETERS
|
host |
|
|
power supply |
AC 220V/50Hz |
|
power |
150W |
|
Working environment temperature |
10℃~35℃ |
|
Working environment humidity |
≤80% |
|
Overall Dimensions |
1000mm (length) x 400mm (width) x 470mm (height) |
|
weight |
70kg |
|
optical system |
|
|
wavelength range |
185nm-900nm |
|
Grating line density |
1800 items |
|
Monochromator |
Czerny-Turner |
|
Spectral bandwidth |
Multi level automatic switching of 0.2nm, 0.4nm, 1.0nm, and 2.0nm |
|
Wavelength indication error |
≤±0.1nm |
|
Wavelength repeatability |
≤±0.025nm |
|
baseline drift |
0.003ABS/30min (static) 0.005ABS/15min (dynamic) |
|
resolution |
bronze lamp324.7nm wavelength, 0.2nm bandwidth, half width within 0.2nm ± 0.01nm |
|
detector |
Hamamatsu photomultiplier tube from Japan |
|
Light source system |
|
|
lamp holder |
Rotating tower type fully automatic switching light rack |
|
Lamp power supply mode |
400Hz, 1/4 duty cycle pulse lights up |
|
Lamp current adjustment range |
(0~10mA (average current) software adjustment |
|
Atomization system |
|
|
Characteristic concentration |
(Cu)0.025μg/mL /1% |
|
detection limit |
(Cu)0.006μg/mL |
|
burner |
0.5mm × 100mm all titanium burner |
|
repetitiveness |
RSD≤0.5% |
|
sprayer |
Metal sheathed glass spray |
|
雾化室 |
Corrosion resistant all plastic atomization chamber |
|
protect |
Multiple automatic protection functions, acetylene leakage alarm, and system shutdown |
|
analytical method |
|
|
measurement method |
air-Acetylene flame method |
|
Concentration calculation method |
Linear regression method automatically fits the working curve and calculates the results automatically |
|
Repeated measurement times |
1~99 times, automatically calculate the average and content, provide standard deviation and relative standard deviation |
|
output result |
Multi format data saving and printing output |
|
communication interface |
standardRS232, USB expansion interface |
|
software environment |
Windows XP/Vista/Windows 7 operating systems, professional Chinese software |
|
Function Expansion |
Hydrogen system and graphite furnace system |