1、 Overview
The ZX119-8 megohmmeter calibration device (with terminal voltage testing) is suitable for testing various models and specifications of domestic and imported insulation resistance meters (megohmmeters) and similar instruments and meters. It can also be used as a DC standard high resistance box, high input internal resistance DC high voltage meter, and grounding resistance meter drag device.
2、 technical indicators
2.1 Indicators of megohmmeter calibration device
|
Step by Step 值Ω |
1011 |
1010 |
109 |
108 |
107 |
106 |
105 |
104 |
103 |
102 |
|
Step disk |
1 |
10 |
10 |
10 |
10 |
10 |
10 |
10 |
10 |
10 |
|
accuracy class |
10 |
5 |
2 |
1 |
0.5 |
0.2 |
0.2 |
0.2 |
0.2 |
0.2 |
|
Rated voltage and rated current |
5000V |
5000V |
5000V |
5000V |
2500V |
1000V |
2mA |
10mA |
20mA |
50mA |
2.2 Indicators of Megohmmeter Terminal Voltage Tester
|
project |
Content |
|
measurement range |
Effective value/peak value from 0 to 1999.9V to 5000V |
|
Maximum allowable error |
±(1%RD+2D) |
|
Input resistance |
10GΩ±10% |
2.3 Usage conditions
Temperature: 20 ℃± 5 ℃
Humidity: ≤ 65% RH
2.4 Storage and transportation conditions
Temperature: -10 ℃~55 ℃
Humidity: ≤ 80% RH
The indicators of the 2.5 megohmmeter calibration device that are not listed meet the requirements of JJG1072-2011.
2.6 Volume and Weight