E-mail
youkewjc@163.com
-
Phone
18601767928
-
Address
No. 228 Qingling Road, Fengxian District, Shanghai
Shanghai Youke Instrument Co., Ltd
youkewjc@163.com
18601767928
No. 228 Qingling Road, Fengxian District, Shanghai
| brand | YOKE/Shanghai Youke | detection limit | 0.01ppm |
|---|---|---|---|
| repetitiveness | . | Measurement range | (0 ~ 20.00)mg/L(ppm) (0 ~ 200.0)%mg/L |
| accuracy | Electricity meter: ± 0.10 mg/L, matching: ± 0.40 mg/Lmg | Measurement principle | Electrochemical probe method - polarographic type |
| price range | 1-5 thousand | Origin category | domestic |
| Instrument type | Laboratory type |
Shanghai Youke DO desktop dissolved oxygen analyzer JPSJ-605F
Shanghai Youke DO desktop dissolved oxygen analyzer JPSJ-605F Instrument features:
1. The dissolved oxygen measurement mode has the functions of automatic temperature compensation, automatic salinity compensation, and automatic pressure compensation.
2. The new structure of the polarographic dissolved oxygen electrode can simultaneously measure temperature, salinity, and dissolved oxygen, and is equipped with a dissolved oxygen electrode calibration kit
Polarization only takes 3-5 minutes.
4. Compliant with international GLP requirements, it has intelligent functions such as automatic calibration, data storage, RS232 output, clock display, and function settings.
5. There are three different measurement modes: stable display mode; Timed measurement mode; Continuous measurement mode.
6. Optional intelligent magnetic stirrer (dissolved oxygen measurement must be used in conjunction with a magnetic stirrer)
Shanghai Youke DO desktop dissolved oxygen analyzer JPSJ-605F Instrument parameters:
Name:Desktop dissolved oxygen analyzer
Model: JPSJ-605F
Measurement range: (0~20.00) mg/L (ppm) (0~200.0)%
Resolution: 0.1/0.01 mg/L (ppm) 1/0.1%
Accuracy: Electrometer: ± 0.10 mg/L, matching: ± 0.40 mg/L
Response time: ≤ 30 seconds (25 ℃, 90% response)
Residual current: ≤ 0.1 mg/L
Temperature compensation range: 0~45 ℃ (automatic)
Salinity compensation range: (0~45) ppt (automatic)
Pressure compensation range: (80~105) kPa (automatic)
Automatic calibration: air saturated with water; Water saturated with air
Electrode type: Polarographic
Measurement range: -10 ℃~110 ℃
Resolution: 0.1 ℃
Accuracy: 5~60 ℃ Range: ± 0.4 ℃ Other ranges: ± 0.8 ℃
Data storage: 600 sets
Storage contents: measurement value number, measurement value, temperature value, measurement date, measurement time
Power supply: DC9V/300mA
Communication interface: RS232
Environmental temperature: 10~30 ℃
Environmental humidity: ≤ 80%
IP rating: IP54 dustproof and splash proof
| Portable dissolved oxygen analyzer P610 | |
Portable Dissolved Oxygen Analyzer P610 Instrument Purpose: |