Phone
13023696116
-
Address
No.11 Shengang Road, Gouzhuang Industrial Park, Hangzhou City
Hangzhou Mingyang Instrument Co., Ltd
13023696116
No.11 Shengang Road, Gouzhuang Industrial Park, Hangzhou City
Salt spray test box price @ News ExpressParameter Specifications
| Product Name | model | Studio size mm | Main technical parameters | Remarks | ||
|
Salt spray corrosion test tryoff | YWX/Q-150 | 400×600×450 |
1.Temperature range: RT~55 ℃
2.Temperature deviation: ± 2 ℃
3.Temperature uniformity: ± 0.5 ℃.
4. Salt spray settling amount: 1-2ml/80 cm2/h. |
1.The box is made of PVC plastic.
2.Tower spraytower.
3.Continuous, intermittent cycle spray.
4. Domestic high-quality productsExecute components.
5. Domestic high-quality productsDigital temperature control instrument.
The customer provides their own air compressor. | ||
| YWX/Q-250 | 600×900×500 | |||||
| YWX/Q-750 | 750×1100×500 | |||||
| YWX/Q-010 | 850×1300×600 | |||||
| YWX/Q-016 | 850×1600×600 | |||||
| YWX/Q-020 | 900×2000×600 | |||||
|
(Export type) Salt spray corrosion test chamber | YWX/Q-150 | 400×600×450 |
1.Temperature range: RT~55 ℃
2.Temperature deviation: ± 2 ℃
3.Temperature uniformity: ± 0.5 ℃.
4. Salt spray settling amount: 1-2ml/80 cm2/h. |
1.The box is made of imported PVC plastic.
2.Tower spraytower.
3.Continuous, intermittent spray
4.Import execution components
5.Fuji temperature control instrument from Japan.
Customer provides their own air compressor | ||
| YWX/Q-250 | 600×900×500 | |||||
| YWX/Q-750 | 750×1100×500 | |||||
| YWX/Q-010 | 850×1300×600 | |||||
| YWX/Q-016 | 850×1600×600 | |||||
| YWX/Q-020 | 900×2000×600 | |||||
| Damp heat salt spray test chamber | SY/Q-750 | 750×1100×500 |
Temperature: 40 ℃, Humidity:≧93%RH The salt spray test parameters are the same as above. | This box can be tested separately for salt spray or wet heat. | ||
| Sulfur dioxide salt spray test chamber | SO2Y/Q-150 | 400×600×450 |
Sulfur dioxide gas concentration: 0.1~1%. The salt spray parameters are the same as above. |
This box can be used as a chemical testing box. 2. This box can be used for both salt spray test and sulfur dioxide test 3. The external steel cylinder method is used to produce sulfur dioxide. 4. Steel cylinders and gases are priced separately. | ||
| Sulfur dioxide salt spray test chamber | SO2Y/Q-250 | 600×900×500 | ||||
| Salt spray wet heat drying cycle box | YSW-1100 | 880×1460×890 |
1.Temperature range: RT~70 ℃
2.Temperature deviation: ± 2 ℃
3.Temperature uniformity: ± 0.5 ℃.
4.Salt spray settling amount: 1-2ml/80 cm2/h.
5.Humidity range: 50%~98% RH
|
1.The box is made of new advanced anti-corrosion materials.
2Tower spray.
3Continuous, intermittent cycle spray.
4Import execution components.
5LCD touch screen programmable temperature and humidity control instrument.
The customer provides their own air compressor. | ||
|
IIIBox material
| |||
| 1 | The test case body is made of light gray PVC board with a thickness of 10mm and a durable temperature of 70 ℃. Surrounded by three-dimensional reinforcement at the edges, the structure is firm | ||
| 2 |
The laboratory cover is sealed with PVC impact resistant board, with a thickness of 5mmHigh temperature use does not distort or deform. Edge inner and outer double
Thickening the layers to prevent distortion and deformation caused by long-term high temperatures. | ||
| 3 | Hidden integrated reagent replenishment bottle, easy to clean and operate | ||
| 4 | The pressure air tank adopts high temperature resistant PC board with excellent insulation effect. (Taiwan Exhibition Electric Heating) | ||
| 5 | The sample placement table is made of 5mm heat-resistant PVC board with strong load-bearing capacity. The test sample rack adopts a flat indexing type, which can adjust the angle arbitrarily, The fog falls evenly on all sides, and the fog is completely consistent. The test results are accurate, and the number of test samples placed is large. (Can be customized according to customer product requirements) | ||
| 4 technical basis | |||
| Operate and set according to JIS, ASTM, CNS, GB standard parameters | |||
| 5、 Salt water spray test: NSS ACSS | |||
| 1 | Laboratory temperature | 35 ℃± 1 ℃ fluctuation: ≤± 1.0 ℃ uniformity: ≤ 2 ℃ | |
| 2 | Temperature of pressure air tank | 47℃±1℃ | |
|
VICorrosion resistance test: CASS
| |||
| 1 | Laboratory temperature | 50 ℃± 1 ℃ fluctuation: ≤± 1.0 ℃ uniformity: ≤ 2 ℃ | |
| 2 | Temperature of pressure air tank | 63℃±1℃ | |
|
7air supply system
| |||
| | The air pressure is adjusted in two stages |
A section for a rough adjustment of 2Kg/cm2The imported air filter comes with a drainage function
| |
|
The second stage requires precise adjustment of 1Kg/cm2,1/4 Pressure gauge, displaying precision and accuracy
| |||
|
8spray method
| |||
| a. Bernoulli principle absorbs salt water and atomizes it uniformly, without any obstruction or crystallization, ensuring continuous testing | |||
| b. The nozzle is made of Taiwan Taiwan floating tempered glass, and the size and angle of spray can be adjusted | |||
|
c. Spray volume 1~2ml/h adjustable (ml/80cm2/The H standard requires testing for 16 hours and taking the average. The measuring cylinder adopts an embedded installation, with a beautiful appearance
| |||
| Neat, easy to observe, reducing instrument installation space | |||
| d. Spray mode: continuous spray | |||
| Nine heating system | |||
| Adopting direct heating method, the heating speed is fast and the standby time is reduced. When the temperature reaches, it automatically switches to a constant temperature state, with precise temperature and low power consumption. | |||
| Pure titanium heating tube, resistant to acid and alkali corrosion, with an ultra long service life | |||
|
TenControl system
| |||
| 1. | The laboratory and pressure tank both use LCD dual character temperature controllers with microcomputer automatic calculation function, with a control error of ± 1.0 ℃ | ||
|
.The circuit board has undergone moisture-proof and anti-corrosion treatment, with high precision and long service life
| |||
| 2. | The laboratory heating tank adopts a liquid expansion safety temperature controller of 30-150 ℃ | ||
| 3. | Digital time controller 1S~99hr (1 unit) | ||
| 4. | Timer 0-999hr | ||
| 5 | Relay (OMRON 1 group) | ||
| 6 | Swing board with light switch, capable of continuous operation 25000 times | ||
|
ElevenWater addition system
| |||
| Adopting both automatic and manual water addition systems, the pressure tank and laboratory water level can be replenished automatically or manually to prevent damage to the instrument due to water shortage and high temperature. | |||
| | |||
|
TwelveMist removal system
| |||
| Clear the salt spray inside the test chamber during the experiment and shutdown to prevent corrosive gases from flowing out and damaging other precision instruments in the laboratory | |||
| Thirteen Safety protection device | |||
|
1.When the water level is low, the power is automatically cut off and the safety warning light device lights up to display
| |||
|
2.When the temperature exceeds the limit, the heater power will be automatically cut off, and the safety warning light device will light up and display
| |||
|
3.When the water level of the test drug (saline) is low, the safety warning light device lights up and displays an alarm prompt
| |||
|
4.Leakage protection function, preventing personal injury and instrument failure caused by circuit leakage or short circuit
| |||
|
FourteenStandard equipment
| |||
| a. Storage rack 1 set | |||
| b. Measuring cylinder... 1 piece | |||
| c. Collector 1 unit | |||
| d. 1 glass nozzle (vulnerable parts need to be replaced regularly) | |||
| e. 1 filter (to be replaced regularly) | |||
| f. Spray tower.................. 1 set | |||
| j. Mist removal system 1 set | |||
| h. Automatic water addition system... 1 set | |||
|
FifteenAttachment
| |||
| Sodium chloride (NaCl) 500g 1 bottle | |||
| 1 copy of operation manual | |||
| 1 5L measuring cup | |||
| 1 salinometer | |||
|
SixteenInstallation environment
1,Require a 2-4 kg air source (with an air pump also available)
2,Power requirement: AC220V 50HZ MAX15A
3,The equipment naturally emits waste salt spray, and it is required to open a ventilation hole on the laboratory wall
| |||