The HDPI-2000B pocket multifunctional process calibrator is a portable instrument with multiple functions and high precision. Specially designed for process control, suitable for industrial field debugging, able to meet a wide range of work task requirements, supporting the measurement and output of almost all process parameters such as voltage, resistance, current, frequency, thermocouple, thermal resistance, etc. It can also calibrate secondary instruments that can measure these parameters with an accuracy of up to 0.05%.
Product Features:
1. Built in one-time quick melting fuse can prevent the majority of instrument damage caused by mistakenly inserting 220V mains power.
2 AD、DA、 High precision and high temperature stability chips are used for important chips such as reference voltage and operational amplifiers to improve the temperature stability and long-term aging performance indicators of the instrument.
3. Reduce the sensitivity of thermal resistance and resistance output to external excitation current in the circuit structure. Improve output accuracy. Suitable for both constant current and non constant current resistance measurement methods, it expands the range of use of the instrument.
4. Faster button response speed with zero latency.
The functions cover the vast majority of maintenance requirements such as PLC, DCS, ESD, field instruments, regulating valves, frequency converters, etc., with high precision and good long-term stability. The security protection is fully in place, the display is clear, the shell adopts branded shell, and the frosted treatment is resistant to dirt and scratches. The contact resistance of all copper connectors is extremely low. Compact size, easy to carry, and reasonable panel layout. V、 mV、 mA 、 Both Ω input and output have corresponding buttons, which can be directly accessed without the need for a combination of buttons, making the operation extremely simple.
Measurement function and scope:
| Function FUNC | output type OUT UNIT |
Set Range | Set Steps SET STEP |
| DC voltage DC. V | V | 0~11.00V | 0.01/0.1/1 |
| Direct Current Millivolt DC.mV | mV | 0~110.00mV | 0.1/1/10 |
| Ohm Ω | Ω | 20~400Ω | 1/10/100 |
| Direct current DC.mA | mA | 0~24.00mA | 0.01/0.1/1/4 |
| Thermocouple TC | R | 0~1700℃ | 1/10/100 |
| S | 0~1600℃ | ||
| B | 500~1800℃ | ||
| K | -200~1370℃ | ||
| E | -200~1000℃ | ||
| J | -200~1200℃ | ||
| T | -200~400℃ | ||
| N | -200~1300℃ | ||
| Thermistor RTD | Pt100 | -200~850℃ | 1/10/100 |
| Cu50 | -50~150℃ | ||
| 24V | 24V mA |
24V: Cannot be set. Current measurement: 0~24.000 mA | none |
| Function FUNC | Measurement type MEASURE UNITS | Measurement Range | Resolution Resolution |
| DC voltage DC. V | V | 0~30.000V | 0.001V |
| Direct Current Millivolt DC.mV | mV | 0~150.00mV | 0.01mV |
| Ohm Ω | Ω | 0~999.9Ω | 0.1Ω |
| Direct current DC.mA | mA | 0~30.000mA | 0.001mA |
| thermocouple TC |
R | 0~1700℃ | 1℃ |
| S | 0~1600℃ | ||
| B | 500~1800℃ | ||
| K | -200~1370℃ | ||
| E | -200~1000℃ | ||
| J | -200~1200℃ | ||
| T | -200~400℃ | ||
| N | -200~1300℃ | ||
| thermal resistor | Pt100 | -200~850℃ | 1℃ |
| Cu50 | -50~150℃ | ||
| Range conversion |
mA | 0~99900 engineering units (applicable to current measurement and 24V output current measurement functions) | Max 4, Max 1 (related to project quantity) |