Phone
13261731718
-
Address
Mingliu Future Building, Majiabao West Road, Fengtai District, Beijing
Beijing Huayi Tongtai Environmental Protection Technology Co., Ltd
13261731718
Mingliu Future Building, Majiabao West Road, Fengtai District, Beijing
BMOZ-2000C Wall mounted Ozone Concentration Online Detector Product Introduction:
The BMOZ-2000C wall mounted ozone concentration online detector is mainly used for detecting ozone concentration at the outlet of ozone generators. It uses ultraviolet lamps as the light source system and has a service life of over 30000 hours. The measurement data is stable and the accuracy is high. Advanced microprocessor control technology is adopted. BMOZ-2000C is a high concentration ozone concentration detector designed based on the characteristics of ozone's absorption of ultraviolet light, according to the Beer Lambert law, and in accordance with the national environmental standard "Determination of Ozone in Environmental Air - Ultraviolet Method" and the calibration method of the Ministry of Construction's "Ozone Generator for Water Treatment".
By measuring the changes in light signal intensity before and after the absorption of ultraviolet light by ozone, the current ozone concentration value can be calculated. The equipment uses an advanced dual path ultraviolet light source system. Equipped with an intelligent management system for light tubes. After turning on, the UV lamp can quickly reach the measurement state. Adopting German separated light cell technology. The equipment has the characteristics of no leakage, high pressure resistance, resistance to high flow sampling gas impact, easy cleaning, convenient maintenance, simple operation, and low usage cost.
The equipment automatically blows the zero point every 2 hours to ensure the stability of the zero point and prevent zero drift from affecting the accuracy of concentration detection. The gas flow rate, pressure, and flow velocity remain unchanged during zero hour calibration. Prevent the impact of airflow on zero calibration. Ensure that the equipment can operate continuously 24 hours a day.
BMOZ-2000C Wall mounted Ozone Concentration Online Detector Product Features:
Measurement principle: According to the Beer Lambert law
Testing method: Dual path UV absorption method with high measurement accuracy
Light source system: Long life ultraviolet lamp tube (wavelength 253.7nm)
Light cell system: Separated light cell technology, leak free, high pressure resistant, and resistant to high flow sampling gas impact;
Intelligent compensation: built-in light source automatic compensation function;
Automatic zeroing: The equipment automatically zeros once every 2 hours to ensure the stability of the zero point and the accuracy of the detected concentration.
Data display: Devin 5.0-inch touch screen, easy for users to operate
Output function: 4-20mA current signal, RS485 communication
BMOZ-2000C Wall mounted Ozone Concentration Online Detector Application Fields:
Ozone generator manufacturers, municipal water industry, industrial wastewater industry, fine chemical industry, food and drinking water industry, space disinfection industry, swimming pool disinfection industry, and other industries that use ozone generators.
Suggested installation and connection diagram for this ozone concentration detector
BMOZ-2000C Wall mounted Ozone Concentration Online Detector Product Parameters:
Range: 0-400 (adjustable) mg/L
Concentration unit: mg/L% (wt) g/NM3 (customizable measurement generator output, unit in kilograms)
Display resolution: 0.01
Accuracy: ± 3%
Output mode: Differential 4-20mA (linear) RS485
Input gas flow rate: 1.0L+0.2L/min
Input gas pressure:<0.1MPa
Power supply: AC 220V+10% 50Hz
Installation size: H: 350 L: 270 W: 130
Gas inlet and outlet pipe diameter: inner diameter 4mm, outer diameter 6mm PTFE tube